Our passive probe when connected to an MSO54 is not reading any voltage even when connected directly to voltage source. Is there any other methods of testing functionality or servicing these probes?

Posted Fri, 30 Sep 2022 19:51:29 GMT by Expert, Tektronix Applications

There is not any documentation regarding functional tests for the TPP1000 or TPP500 passive probes. Could it be that the probe is connected to a earth-isolated voltage source? Without a ground reference to the source you won't get a signal because the scope itself is earth grounded.