Posted Tue, 11 Jul 2023 01:20:27 GMT by Rive, Jorge
I am able to trigger the DAQ6500 from the SMU 2470.  I can plot the samples in the graph view in each of the units.  However, I can't seem to store the samples in the reading buffers for export (in either unit)..  This seems to work fine when I use the units individually, but not when I'm going through the  trigger-flow.  The buffers are empty (even though data is correctly sampled, since I can plot it in the graph views) - where is the data going-?--the assigned reading buffer in each unit remain empty - but, again, data is being sampled correctly (it seems) since the graphing view in the respective units displays them without any issue.  
Posted Wed, 12 Jul 2023 16:38:48 GMT by C, Andrea

On the graph view, where you have data, what is name of the Active buffer.  Go to Data tab once Graph view is displayed.

Do your trigger blocks make use of same buffer name?

Posted Wed, 12 Jul 2023 16:44:51 GMT by C, Andrea
On the graph view, where you have data, what is name of the Active buffer.  Go to Data tab once Graph view is displayed.

Do your trigger blocks make use of same buffer name?


Trigger Block using Buffer for Measurements
Posted Thu, 13 Jul 2023 13:08:08 GMT by Rive, Jorge

I figured it out...I needed to set the output to Reading Row to get all the entries in the buffer out.

Thank you for responding.

Jorge