Posted Wed, 24 Jan 2024 04:38:14 GMT by REYES, KIMBERSON
Hello,

Currently we are using DMM 2750 to measure resistance values. After measurement, we manually record the result in a paper and then input again the values into excel to make test report.

I would like to ask if it is possible to automate the process of recording the measured values? then export the data into a computer?
If it is possible, What are the requirements that we need to prepare in order to implement this automation?
How will be the set up and does it need special programming from an expert?

Regards to all & Thank you very much in advance.
Posted Mon, 29 Jan 2024 02:25:11 GMT by C, Andrea
Are you using any scan cards with the 2750?  Or using it as a DMM to measure Ohms of a single devices at a time.

The 2750 can be controlled by sending the SCPI commands that it understands over GPIB or RS-232.
The SCPI commands are documented in the 2750 manual.

The KickStart software can interface with 2750 over GPIB interface.
The Windows computer will need a GPIB interface from National Instruments or Keithley (KUSB-488B).

If using scan cards, use the datalogger application of KickStart (KICKSTARTFL-DL).
If not using scan cards, use the DMM application of KickStart (KICKSTARTFL-DMM)

More info:KickStart
Posted Thu, 22 Feb 2024 04:54:45 GMT by REYES, KIMBERSON
Hello Andrea,

Thank you for your answer.

We are using DMM 2750 to measure Ohms of a single devices at a time.
Is it possible that we just use GPIB connection & ExceLINX software so that the 
results are automatically recorded in a worksheet?

Thank you.