Posted 13 days ago by junchi, Song
I'd like to do delta-mode measurements with the Keithley 6221 & 2182A devices.
I have the 6221 and 2182A connected through RS232 null modem and trigger link cable, and the 6221 is connected to PC through GPIB bus. When I  configure the mode( 19.2k rate, Xon_Xoff flow control, CR terminal),and take the button "DELTA" in the GPIB interface, 6221 reports error-241"hardware missing". I can't find the problem.
 
Posted 13 days ago by C, Andrea
Maybe obvious, but this error means the 6221 does not receive the expected response from 2182A on the rs-232.

Double check the rs-232 settings are same in both instruments.
Then set LAN or GPIB the active interface on 6221.
Press delta button on 6221 and accept defaults.
Error?
Posted 11 days ago by junchi, Song
Thanks for your answer, I have checked this for several times. Now I am trying to renew the RS232 cable,because I didn't use the original assembly line before. 
 