You’re offline. This is a read only version of the page.
Skip to main content
Tektronix, Inc.
Toggle navigation
Products
Solutions
Calibration & Services
Support
Support
TekTalk
All
All
Forums
Search Filter
Forums
Enter your search term. Use asterisk (*) as a wildcard character.
English
English
Việt Nam
русский
中文(台灣)
中文(中国)
日本語
한국어
français
Deutsch
Login
Home
TekTalk
Oscilloscopes
TDS 2024B Fails Adjustment Procedure
TDS 2024B Fails Adjustment Procedure
Posted
Tue, 13 Dec 2022 22:31:59 GMT
by
Kay, Jon
I have a TDS 2024B that I am trying to run through the adjustment procedure. I'm using a Fluke 9500B O'scope Calibrator. It runs through the DC voltages on channels 1 to 4 but fails when I get to the -2.5V step on EXT TRIG. This is step 33 of the procedure. Anybody have a possible solution or cause for this failure? I would appreciate any ideas.
Thank you very much.
Posted
Wed, 14 Dec 2022 00:18:38 GMT
by
Teles, Afonso
Hi,
The
TDS2000B service manual
lists step 33 (page 5-7) as 1.0 V, not -2.5 V.
Posted
Wed, 14 Dec 2022 15:39:44 GMT
by
Kay, Jon
Thank you very much.... I had the manuel for the TPS model not the TDS.....very embarassing.
Thanks again.
You must be signed in to post in this forum.