Posted Tue, 16 Jan 2024 19:02:36 GMT by Teles, Afonso
Hi Makis,

If you notice, when your signal is outside of the screen, you will get clipping warning on the channel tab and measurement badge. This is because you're overdriving the ADC and any captured signal is invalid.
Additionally, any measurement taken is essentially undefined behavior.

No, the calculation doesn't depend on vertical scale (except for unavoidable side effects like differing noise, resolution, etc)
Posted Tue, 16 Jan 2024 21:47:24 GMT by Teles, Afonso
Hi Makis,

From my quick testing, neither of those happens when a clipped measurement is taken. Instead, it clears out the whole histogram and doesn't start rebuilding it until the channel stops clipping.
Posted Fri, 19 Jan 2024 12:36:45 GMT by Valderanis, Chrysostomos Research Associate
Hi Alfoso,

well, in my testing the histogram keeps going on. I can never make it to clean itself (without changing parameters, that is).

Cheers,
Makis
 
Posted Tue, 23 Jan 2024 00:07:25 GMT by Teles, Afonso
Hi Makis,

Please make sure you're on the latest version.
If this issue remains, I'd ask if you could share a setup file.

You must be signed in to post in this forum.