Posted Fri, 10 Feb 2023 18:28:31 GMT by Li, Xiang
It's almost impossible to get integer number of cycles on the scope so how do you collect duty cycle stat on the badge? Do you throw out the partial cycles on the left/right most and only collect from the full cycles? 

For the attached screenshot: would it be collecting stats from those 5 pulses? 
Posted Mon, 13 Feb 2023 17:38:31 GMT by Teles, Afonso
Hi Xiang,

Yes the algorithm is smart enough to exclude partial cycles and only collect duty cycle data on full cycles.

If you put the instrument into Single mode and look at N on the stats badge, you will be able to see the number of cycles the algorithm has detected in the captured waveform and is doing duty cycle measurements on.

You must be signed in to post in this forum.