Posted Wed, 22 Feb 2023 13:47:31 GMT by Wade, Jay
I'm having a hard time trying to locate the specs of Option 10 for the RSA3408B spectrum analyzer. If anyone has a link or a PDF to share, that would be greatly appreciated.


Thanks!
Posted Wed, 22 Feb 2023 23:25:27 GMT by Teles, Afonso
Hi Jay,

The specifications of the measurements possible with Option 10 should match the specifications of the hardware, there's no need for a separate spec for that option.

If you have a certain measurement spec that you need I can give you some more specific help in finding/calculating it.
Posted Thu, 23 Feb 2023 13:22:38 GMT by Wade, Jay
Hello,

   So I'm curious if the THD measurement is restricted to only the audio spectrum (30Hz to 20kHz), or does it apply to the full spec of the analyzer? (DC to 8GHz)


Thanks!

Jay
Posted Fri, 24 Feb 2023 15:43:21 GMT by W2AEW, Alan Sr. Field Applications Engineer
The THD measurement can be made on audio signals - either directly (baseband), or as demodulated from an AM or FM modulated carrier.  The measurement can be performed on audio frequencies from 20Hz to 100kHz.

You must be signed in to post in this forum.