You’re offline. This is a read only version of the page.
Skip to main content
Tektronix, Inc.
Toggle navigation
Products
Solutions
Calibration & Services
Support
Support
TekTalk
All
All
Forums
Search Filter
Forums
Enter your search term. Use asterisk (*) as a wildcard character.
English
English
Việt Nam
русский
中文(台灣)
中文(中国)
日本語
한국어
français
Deutsch
Login
Home
TekTalk
Semiconductor Parameter Analyzers
Difference in UTM time output data Clarius vs. KXCI, decimal place missing
Difference in UTM time output data Clarius vs. KXCI, decimal place missing
PLEASE BE ADVISED: The content of this posting may contain high-voltage safety risk that could potentially result in fatality. The answers provided by Tektronix experts are specific to the question posed in the Thread, and should not be applied to all situations, regardless of similar issues, products, or conditions. Please refer to the safety section of your product manual.
Posted
about a year ago
by
H., Viktor
Scientist
Hello, we are working with our own pulse measurement UTM, that is based on seg_arb commands. Strangely, the KXCI output data differ from the Clarius output data when running the same UTM with identical input parameters. If the total sequence length exceeds 1ms, the time values from KXCI have only 7 decimal places whereas in Clarius we get 8 decimal places. For shorter sequences the data is identical. The voltage and Current values seem to be unaffected. Tested in Clarius v. 1.6.1 and 1.11 with identical behavior.
Is there any explanation for this inconsistency? Any idea how to fix this would be highly appreciated.
Thank you.
Posted
8 months ago
by
C, Andrea
Tek Expert
In version 1.13 of our Clarius software, KXCI API is enhanced for the PMU and accessing the SegArb mode directly without UL and EX to KULT code.
Numerous examples have been posted to our github site:
Github 4200A KXCI Python Samples
You must be signed in to post in this forum.