Difference in UTM time output data Clarius vs. KXCI, decimal place missing
Difference in UTM time output data Clarius vs. KXCI, decimal place missing
Posted
Wed, 31 Jan 2024 15:23:28 GMT
by
H., Viktor
Scientist
Hello, we are working with our own pulse measurement UTM, that is based on seg_arb commands. Strangely, the KXCI output data differ from the Clarius output data when running the same UTM with identical input parameters. If the total sequence length exceeds 1ms, the time values from KXCI have only 7 decimal places whereas in Clarius we get 8 decimal places. For shorter sequences the data is identical. The voltage and Current values seem to be unaffected. Tested in Clarius v. 1.6.1 and 1.11 with identical behavior.
Is there any explanation for this inconsistency? Any idea how to fix this would be highly appreciated.
Thank you.
Posted
Fri, 26 Jul 2024 11:59:08 GMT
by
C, Andrea
In version 1.13 of our Clarius software, KXCI API is enhanced for the PMU and accessing the SegArb mode directly without UL and EX to KULT code.
Numerous examples have been posted to our github site:
Github 4200A KXCI Python Samples