Posted Wed, 28 Aug 2024 09:39:10 GMT by Sharma, Vipin
My AFG 31022 is not switching to sequence run mode from continuous run mode using screen touch panel and even with buttons . The options after ch1 WFM and Ch2 WFM i.e options namely Repeat, Wait Event, Jump Event , Jump addr
etc. are greyed out. Please suggest me some solution
Posted Thu, 29 Aug 2024 21:41:23 GMT by C, Andrea
The SEQ does have separate license to have all features such as repeat, jump, etc.
Without the license, can load one waveform and continuous mode.

Obtain Trial License
Posted Fri, 30 Aug 2024 02:37:23 GMT by Sharma, Vipin

Hi, Thanks for the information. I have a file which contains 13M points and I wish to upload it in AFG31022 using ArbExpress but it shows the "error code 1605". Going by the manual of AFG31022, it says that we can upload upto 16 M points in continuous running mode in advance section. However it is not even accepting 1 M points. Please suggest a solution.

You must be signed in to post in this forum.