Hello,

Regarding the "Measure Count" in the LabVIEW driver KE26XX Configure Measure Settings.vi, since the default Measure Count to 1, this means that the SMU, for example, makes one Current measurement reading at every Voltage value, doesn't it?

Also, if I set the Measure Count in the vi to 10, will the SMU give an average measured Current value from the 10 measurements it made at every Voltage value?
Posted Sat, 13 Jul 2024 12:07:47 GMT by C, Andrea
Measure count of 10 would give 10 individual readings in the buffer.
To obtain an average, you could use a digital filter.  The instrument has both moving and repeating average filter.
Or there are buffer statistics that you can query to obtain mean, std dev, etc.