Posted Thu, 21 Nov 2024 13:05:11 GMT by Nawaz, Hamza

Is it possible via kickstart to close just the hi or low relays in the 7702 expansion card using the daq6510?

example:

CH1 High - 5V

CH1 Low - GND1

CH2 High - 12V

CH2 Low - GND2

Is it possible to take an array of measurments ex:

CH1 High to CH2 High

CH1 High to CH2 Low

CH2 High to CH1 Low

Posted Wed, 18 Dec 2024 13:06:17 GMT by C, Andrea
KickStart does not offer relay control of that nature.  It supports conventional scanning of the two-pole channels on the multiplexer cards.

The internal web page has an application that provides a means to send commands to the instrument over LAN.
Our use our TSP Toolkit from vscode to send commands.

To do what you describe would require connection of the front terminal DMM to the 7702 card input or card sense so that you can measure between the two banks.  When wiring the channels, treat them as single pole and use only the HI or only the LO.
NOTE:  you will also have to manage channel 43 for the card sense to bank 2.

 