Posted Mon, 20 Nov 2023 11:55:48 GMT by Parthenios, John
I'm trying to set up a van der Pauw method for mesuring resistivity of 2D material films. I have only an SMU with 2 channels (2604B). Is this feasible?
 
Posted Thu, 23 Nov 2023 13:28:02 GMT by C, Andrea
With only two SMU channels, you will have to move cables for each step of the VDP measurement.
But for each step, one SMU forces current and other SMU forces 0A and measures V.

Is easier if you have 4 SMU channels, one for each contact point in the VDP test structure.
As needed for each step, each SMU can be current source, LO return or voltage measure, so no need to move cables.

You must be signed in to post in this forum.