Posted Wed, 19 Jun 2024 07:31:03 GMT by Sagmeister, Michael
Hi folks,

hope the question is not too dumb...

I am trying to understand how the scan list exactly works, as we had some issue (could also be hw related...)

scan.create(1001:1010,"dcvolts")

Easy at it is...

how smart is the Keithley 3700A?
  • will it detect stuck relays?
  • will it wait till relays "bouncing" is "finished"?
  • does it really work like "close 1001 --> measure --> open 1001 --> close 1002 --> measure --> open 1002..."?

Thanks in advance!

br
Michael
Posted Sat, 13 Jul 2024 12:18:23 GMT by C, Andrea
Depending on the connection rule in use, when moving to next channel in a scan, the opening of prior channel and closing of next can occur simultaneously or can be configured for break before make, etc.

In general the scan is just closing, measure, open/close next channel, measure, etc.

Since a measure function is associated with the scan, the instrument will also manage the backplane relays to connect the switch to the internal DMM.

You must be signed in to post in this forum.