The +ILK is shorted to -ILK with a physical wire.
If using one of our 37xx-ST screw terminal accessories, we make this connection for you.
But if custom wiring to the DBxx mass termination connectors, keep in mind you need to either wire them to a interlock switch or jumper them if you want to close backplane relays on the 37xx card.
For just closing channels, either send a list of channels to be closed, or create a pattern and close/open by pattern name.
- -- some TSP commands for channel closure
-
- reset()
-
- channel.close("1001,1911")
-
- print("******************")
- print(channel.getclose("allslots"))
- print("******************")
-
-
- channel.open("allslots")
-
- channel.pattern.setimage("1001,1911","patternA")
-
- channel.close("patternA")
- print("******************")
- print(channel.getclose("allslots"))
- print("******************")
-
- channel.pattern.delete("patternA")