I removed Q421 (current regulator) and some features of channel 1 started to work.
I can control the output voltage but offset and scaling are wrong. At 0V set, the output voltage is 3.7V
and at 15V set, the output is 13.6V. Output voltage measurement are incorrect. At +3.7V, the measured
voltage is -6.3V, at 13.6V it's -22.7V.
Current limit, current measurement and output impedance control work well.
I'm guessing the calibration coefficients are wrong.
I tried to calibrate channel 1 and the calibration procedure seems to be passing
but I can't complete the calibration since channel 2 isn't working.
Could you let me know the part number of Q421 please? Is it the 1N5297UR?