In the manual, it lists the MSO44 as being able to do 500 uV/Div. When using the TPP1000 probes or the TAP1500, the oscilloscope will not go below 5mV/Div. Is there something I am doing wrong, or is it a probe limitation?
Hello,
Both of these are 10X attenuation probes, so the scope needs to multiply the voltage at the interface by 10. This is why you see a 5mV/Div minimum. To avoid this, you would need to use a 1X probe.