Posted Sat, 18 Mar 2023 05:54:54 GMT by Wang, Frank
I noticed that under the measurement for rise time on the MSO 4 series, there is an option to use clock recovery.  Can you provide an example of when I would use this option?
Posted Mon, 20 Mar 2023 17:49:41 GMT by Teles, Afonso
Hi Frank,<br> <br> The clock recovery option is meant for testing the rise time of a serial bus, limiting the measurement to transition bits.
Posted Mon, 20 Mar 2023 23:56:40 GMT by Wang, Frank
<p>Thanks for the reply, but I'm still a bit confused.&#160; Wouldn't all the transitions on the data line be transition bits and thus not need the clock?&#160; Can you provide a concrete&#160;example of when clock recovery would be needed?</p> <p></p>
Posted Tue, 28 Mar 2023 23:37:15 GMT by Teles, Afonso
Transition bits are those which go from one digital state to the other and then back in the least amount of clock cycles possible (2 clock cycles).<br> <br> As an example, for rise time, the transition bits would be those that meet a 1 0 1 pattern.<br> A 1 0 0 1 pattern would not meet this definition and so would be excluded from the measurements. 1 0 1 1 would as that meets the 1 0 1 pattern (what comes after does not matter).<br> <br> These transition bits are often the most testing (aka worst case) of a system and so there's particular interest in measuring them.