You’re offline. This is a read only version of the page.
Skip to main content
Tektronix, Inc.
Toggle navigation
Products
Solutions
Calibration & Services
Support
Support
TekTalk
All
All
Forums
Search Filter
Forums
Enter your search term. Use asterisk (*) as a wildcard character.
English
English
Việt Nam
русский
中文(台灣)
中文(中国)
日本語
한국어
français
Deutsch
Login
Home
TekTalk
Signal Generators (AWGs, AFGs & VSGs)
AFG 1022 Sweep function does not appear to work in linear mode
AFG 1022 Sweep function does not appear to work in linear mode
Posted
Wed, 07 Jun 2023 15:52:49 GMT
by
Johnson, Michael
I'm using an AFG 1022 running version 1.3.0.<br> <br> I setup channel one to drive a Tek scope to verify the output of the AFG 1022.<br> <br> When setting up the AFG 1022 to output a sweep function in log mode, the Tek scope shows the sweeping frequency changing as expected.<br> <br> However, when setting up the AFG 1022 to output a sweep function in linear mode, I do not anything but a steady dc state on the scope.<br> I have tried setting the trigger mode internal, external and manual without success.<br> <br> Has anyone successfully used the AF 1022 using the sweep function in linear mode?<br> <br>  
Posted
Wed, 07 Jun 2023 21:31:44 GMT
by
Teles, Afonso
Hi Michael,<br> <br> I don't see a 1.3.0 available? The latest firmware available on our website is 1.2.7.<br> <br> With 1.2.7 I was able to get a linear sweep with no issues.
Posted
Thu, 08 Jun 2023 15:09:03 GMT
by
Johnson, Michael
<p>1.3.0 was available in May of 2022. It is no longer listed on the website.</p> <p></p> <p>I have attached the 1.3.0 upgrade file. I'm trying to downgrade to 1.2.7 and getting a failure code 6</p> <p>I'm trying to figure out what the error code means.</p> <p>Perhaps you could try the 1.3.0 upgrade to see if you experience what I have re: sweeping in linear mode.</p> <p></p>
Posted
Thu, 08 Jun 2023 15:27:26 GMT
by
Johnson, Michael
<p>I found this from an article on the web regarding error codes via the upgrade process.</p> <p></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">Error message</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">2 File size too large</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">3 File read error</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">4 CRC (Cyclic Redundancy Check) error</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">5 Type flag error in the firmware file</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">6 The firmware version of the instrument cannot be updated to the firmware version of the file.</span></span></span></span></p> <p><span style="font-size:11pt;"><span style="line-height:normal;"><span style="font-family:"Calibri",sans-serif;"><span style="font-family:"Arial",sans-serif;">7 The instrument model does not match with the model of the firmware file</span></span></span></span></p> <div style="left:800.059px;top:969.835px;font-size:14.5298px;font-family:sans-serif;"></div> <div style="left:800.059px;top:969.835px;font-size:14.5298px;font-family:sans-serif;">The error message 6 seems to indicate that "down grading" is not possible. If this is true, is there a process to reset the firmware completely and then do an upgrade?</div> <div style="left:800.059px;top:969.835px;font-size:14.5298px;font-family:sans-serif;"></div>
Answer
Posted
Fri, 09 Jun 2023 16:13:51 GMT
by
Teles, Afonso
<p>Hi Michael,</p> <p></p> <p>I have confirmed that this is an issue with 1.30 and that it's not possible to downgrade the firmware.</p> <p>Our teams are working on an update to fix this issue, I apologize for the inconvenience.</p>
Posted
Fri, 09 Jun 2023 17:30:41 GMT
by
Johnson, Michael
Ok Thank you
Posted
Tue, 27 Jun 2023 16:44:06 GMT
by
Johnson, Michael
Has there been any progress on this issue?
Posted
Fri, 30 Jun 2023 17:46:47 GMT
by
Teles, Afonso
Hi Michael,
I do not yet have any news on a software fix for this issue.
I am working with our teams on other possible solutions and I will contact you through email to discuss them.
You must be signed in to post in this forum.