Posted Wed, 13 Mar 2024 14:49:15 GMT by ochoa, jonathan
Good afternoon, I would need your assistance in finding a solution to connect the TDS 3052 oscilloscope to a PC running Windows 8 using TekScope Utility. I have downloaded version 1.4.2 of TekScope Utility, but so far, I have not been successful in setting it up. I have installed a PCI card for RS232 ports on my PC, and I have also tried using an RS232 to USB converter. Unfortunately, I have not yet achieved success. Any guidance or help you can provide would be greatly appreciated. Thank you in advance.
Posted Wed, 13 Mar 2024 17:25:55 GMT by Teles, Afonso
Hi Jonathan,

How's your RS-232 connection setup on the scope and in VISA? Make sure they share the same settings (baud rate, flagging, parity, etc).
Also make sure that your VISA is actually searching for RS-232 devices (TekVISA doesn't by default).
Posted Thu, 14 Mar 2024 19:59:58 GMT by ochoa, jonathan
Hello Afonso Teles,
 thank you very much for your answer, unfortunately I have not been able to connect it to the pc, it could be something of fireware? the oscilloscope has version 2.11. any suggestion will be taken into account and thank you very much for your time.
Regards!
Posted Fri, 15 Mar 2024 20:51:01 GMT by Teles, Afonso

Hi Jonathan,

It shouldn't be a firmware version problem, but here's the latest version if you wish to upgrade: https://www.tek.com/en/support/software/firmware/tds3000-and-tds3000b-series-firmware-fv341

Pages 3-64 through 3-66 discuss configuring and troubleshooting the RS-232 connection, I would suggest starting there.