I am looking for information on how to do your own calibration and update the internal table on the RSA306B. I would like the amplitude of the unit to be at 0 dB.

Posted Fri, 30 Sep 2022 19:51:21 GMT by Expert, Tektronix Applications

Thanks for contacting us,

There is not a mechanism for self-calibration of the RSA306B. While you can follow the procedures in the Performance Verification manual (https://download.tek.com/manual/RSA306B-Real-Time-Spectrum-Analyzer-Specifications-Performance-Verification-077124301.pdf) to characterize the instrument and verify it is within spec, if the results indicate it needs to be recalibrated it will need to be sent to a calibration lab that supports it. You can find more detail on our calibration services webpage here: https://www.tek.com/en/services/calibration-services