Posted Wed, 20 Mar 2024 18:45:04 GMT by S, Roy
I'm using two SMUs on a 4-terminal DUT.  I'd like to have the two SMUs, which are both sourcing voltage and measuring current, to have their Force Lows be earth ground. When I connect a lug cable to the chassis in the back, I see that I'm connecting earth ground to the shield of the force low terminal, but not the pin/inner conductor of the force low terminal. 

My goal is to have those force low inner conductors both sinking into ground. How do I go about that?

Thank you. 
Posted Sat, 23 Mar 2024 13:54:43 GMT by C, Andrea
Sounds like you have a model 2450 SourceMeter.
You could use a triax shorting plug on the Force LO to connect the center pin to outer shell which is at chassis ground.
Keithley part number for shorting plug:  CS-1546
Info about CS-1546 Triax Shorting Plug

You must be signed in to post in this forum.