Posted Thu, 21 Mar 2024 13:14:33 GMT by Boggs, Christopher
While writing some code for automatically testing duty cycle dependent on pulse width trigger, I found that inside the MSO4000 and DPO4000 Series Programmer Manual on page 2-473 under `TRIGger:A:PULSEWidth:WHEn` the keyword to set the trigger when greater than option is missing. Ran a query on the scope and found out the keywork is supposed to be more on my DPO 4054. I believe the `than` keyword that is currently listed should have been `more` since I don't see a need for the keyword `than`. Please update the documentation so no one else falls for this :)
Posted Thu, 21 Mar 2024 14:29:21 GMT by Wilhelm, Nick
Thank you for reaching out! You are correct, the current PULSEWidth:WHEn argument for greater than is not valid. I have updated the manual to change the argument value from "than" to "MOREthan" (see attached PDF). This updated PDF will be on tek.com later today. 

Nick Wilhelm
Customer Documentation