Using a TDS3054B scope. Manual says" 2%" accuracy. Is this on the reading or the range? Like if I am using 5mV per division, is the error 8div*5mV/div*2%?.

Posted Thu, 29 Sep 2022 23:14:20 GMT by Expert, Tektronix Applications

We specify DC Gain Accuracy as of the reading. For example if you are looking at a 50mV reading you can expect it to be within ±1mV.