Posted Thu, 07 Dec 2023 16:52:11 GMT by Rowe, Martin Other
I'm trying to use TekScope software with MSO46 over USB. It will not connect. The dialog box says None for USB device. TekVISA is installed. When I connect or disconnect the USB cable, A windows pops up saying that a resources is no longer available, then another saying new resources when cable reconnected.
 
Posted Thu, 07 Dec 2023 19:44:33 GMT by Xu, Iris
Hi Martin,
Copying over the reply to the ticket that you sent in:
Are there any other VISAs downloaded to your PC? There is this bug where TekVISA needs to be the first VISA installed if you're using others (like NI-VISA). Does the MSO46 show up in TekVISA, even if not in TekScope?
 
Does the scope show up at all in the windows Device Manager (even if it isn't labelled as MSO46B)? Maybe try a few different USB cables as well, preferably USB 3.0 with a ferrite bead.
 
We will start from here and try to narrow down where the issue is occurring.

Best,
Iris Xu
 
Posted Thu, 07 Dec 2023 20:22:31 GMT by Rowe, Martin Other

<p>&amp;#160;Device Manager shows USB Test and measurement Device IVI so that's good. Checked IO setting of scope. USB device port is on. TekScope still says scope is offline. Tried connecting again. TekScope says Unable to connect to remote scope.</p> <p>No other versions of VISA installed</p>

Actually, that may not be true. I have run NI VistrualBench, which installed software on its own. It's possible that NIVISA is running. S earch of my PC did not turn up NI VISA but who knows?

Posted Fri, 01 Nov 2024 10:05:10 GMT by Zheng, Haolin
Hi Xu,

I also have the same problem with MSO46. With TekVISA installed, I can write command to the scope but have error when read feedback from the scope. (so USB cable should not be the issue) TekVISA is the only VISA installed.

Do you have other suggestion?
 