Digital Multimeters
Problem Using DMM6500 with 2000-Scan Card
Problem Using DMM6500 with 2000-Scan Card
Posted
Wed, 09 Nov 2022 13:06:04 GMT
by
Weiss, Alexander
Prüfingenieur
Hi,I have a DMM6500 and a 2000 scan card. I would like to perform a 2-wire measurement on channels 1 and 2.
The Scan Card is also recognized and displayed correctly on the multimeter and it can also be measured. If I now want to address the ScanCard via:
driver.ChannelFunction(ChannelList:="1:4") = MeasureFunction(ComboBox1.SelectedItem.ToString())
i receive:
:ex. Message = "KeithleyDMM6500: Instrument model does not support this feature: put_ChannelFunction."
I use the vb.net example MeasureMultipleVBnet as a basis. I currently found no way to address the individual channels here. Are there any examples here?
Posted
Thu, 10 Nov 2022 16:43:05 GMT
by
McKinney, Ty
Hello,
I do not believe the driver supports scanning with the DMM6500. The example you're referencing states it's a DAQ6510 example, which would explain why the error is saying that the command is not supported by the DMM6500. See attached picture for DAQ ONLY capable commands.
Posted
Mon, 14 Nov 2022 17:31:43 GMT
by
C, Andrea
Looks like this one found a work around:
https://forum.tek.com/viewtopic.php?f=617&t=142379
Posted
Thu, 24 Nov 2022 19:39:05 GMT
by
Weiss, Alexander
Prüfingenieur
Okay, Thanks, I'll test right away when I'm back in the office
You must be signed in to post in this forum.