Hello my friends,
I 'm doing test on wafer with CV measurement. As requirement testing, the result should be with in 3 sigma ( below 1%).
But actually, Almost result is upper 1%, this problem only happen at frequency 1Khz. Frequency upper than 1Khz almost result is below 1% ( within 3-sigma)
Below is prameter config and result.
Have anyone testing like that or have exprerience about this testing, Please give me advise, thanks so much
Config |
preSoak Voltage |
Hold Time (s) |
Conidtion Test |
Average(pF) |
Standard Deviation |
3Sigma |
- DC Bias : 0.3v
- AC Drive Signal: 30mV
- Interval: 20
- Mode: Sampling
- Freq: 1kHz |
-0.3 |
10 |
1 |
1.424814 |
0.040005262 |
8.42% |
30 |
2 |
1.401139 |
0.033187129 |
7.11% |
60 |
3 |
1.4367645 |
0.041729935 |
8.71% |
-1.5 |
10 |
4 |
1.8695225 |
0.066384102 |
10.65% |
30 |
5 |
1.886752 |
0.04348203 |
6.91% |
60 |
6 |
1.882425 |
0.042995434 |
6.85% |
1.5 |
10 |
7 |
1.896749 |
0.047060789 |
7.44% |
30 |
8 |
1.9012545 |
0.042947112 |
6.78% |
60 |
9 |
1.907959 |
0.053639368 |
8.43% |
3 |
10 |
10 |
1.4861455 |
0.063937909 |
12.91% |
30 |
11 |
1.4844695 |
0.049281225 |
9.96% |
60 |
12 |
1.5566555 |
0.057454766 |
11.07% |